Document code

Unit

HygroFlex HF8 Humidity Temperature
Transmitter: User Guide

Instruction Manual

Document Type

Page 18 of 35

Document title

© 2012; Rotronic AG E-M-HF8-V2_10


Terminal block K3 (RS-485): , terminals K3-1 and K3-2 can be used to power the HF8 from a 15 to 24 VDC
power supply connected to the RS-485 main data line. In this case, do not use terminals K6-1 and K6-2
(normally used to power the HF8).


K3-2: GND

VDC (-), when HF8 is powered from RS-485 data cable

K3-3: D+

RS-485 +

K3-4: D-

RS-485 -

Terminal Block K4 (Relay 3)

Description

K4-1: NC

Normally closed (relay not energized)

K4-2: COM

Common

K4-3: NO

Normally open (relay not energized)

Terminal Block K5 (Relay 4)

Description

K5-1: NC

Normally closed (relay not energized)

K5-2: COM

Common

K5-3: NO

Normally open (relay not energized)

Terminal Block K6 (power Supply)

Description

K6-1: -

Protective ground

K6-2: +

Power supply: VDC (+) or VAC (Phase)

K6-3: -

Power supply: VDC (-) or VAC (neutral) see also 6.4.3

Terminal Block K7 (Relay 1)

Description

K7-1: NC

Normally closed (relay not energized)

K7-2: COM

Common

K7-3: NO

Normally open (relay not energized)