E-m-hf8-v2_10 – ROTRONIC HF8 User Manual
Page 24
E-M-HF8-V2_10
Rotronic AG
Bassersdorf, Switzerland
Document code
Unit
HygroFlex HF8 Humidity Temperature
Transmitter: User Guide
Instruction Manual
Document Type
Page 24 of 35
Document title
© 2012; Rotronic AG E-M-HF8-V2_10
Main Menu
Menu Items
Submenu
Items
Selections or
Information
Notes
Probe 1
(applies to
HygroClip 2
probe only)
Info
Serial Nbr
Version
Firmware version
Name
User defined (HW4)
Humi Adjust
Acquired Points
Lists the cal. points
present in the probe
memory (a max. of
10 points are
shown)
Additional Options:
o
Clear all cal. points
o
Clear the last point
Acquire
( Ref = Probe)
Saves value measured by
the other probe as a cal.
point
Acquire
(Enter Ref)
Manual entry:
known reference
environment
Saves value entered
manually as a cal. point
Adjust
Adjusts the probe
Effect depends on number
of calibration points in
probe memory
Reset to
Factory
Returns the probe to the
initial factory adjustment
Temp Adjust
Acquired Points
Lists the cal. points
present in the probe
memory (a max. of 2
points are shown)
Additional Options:
o
Clear all cal. points
o
Clear the last point
Acquire
( Ref = Probe)
Saves value measured by
the other probe as a cal.
point
Acquire
(Enter Ref)
Manual entry:
known reference
environment
Saves value entered
manually as a cal. point
Adjust
Adjusts the probe
Effect depends on number
of calibration points in
probe memory
Reset to
Factory
Returns the probe to the
original factory adjustment
Probe 2
(applies to
HygroClip 2
probe only)
See Probe 1
See Probe 1
See Probe 1