beautypg.com

Accelerometer-only calibration – PNI SeaTRAX User Manual

Page 23

background image

PNI Sensor Corporation

DOC#1018154 r02

SeaTRAX User Manual

Page 18

gravitational vector, which in turn will result in an inaccurate heading reading. For this

reason, the SeaTRAX should be stationary when taking a measurement.

As previously mentioned, PNI calibrates the accelerometer in its factory prior to shipment.

But over time the bias and offset of the accelerometer will drift. For this reason PNI

recommends the accelerometer be recalibrated every 6 to 12 months. The user may return

the SeaTRAX to PNI for accelerometer calibration, or the user may perform a user

accelerometer calibration. The remainder of this section covers the user accelerometer

calibration.

5.2.1

Accelerometer-Only Calibration

The requirements for a good user accelerometer calibration differ significantly from the

requirements for a good magnetic calibration. Specifically, a good accelerometer

calibration involves the SeaTRAX experiencing a wide range of pitch and roll values,

preferably seeing both ±180° of pitch and ±180° of roll. Also, it is necessary for the

SeaTRAX to be very still during an accelerometer calibration. If possible, PNI

recommends using a fixture to hold the device during calibration, although resting the

SeaTRAX on a hard surface normally is sufficient.

The accelerometer either can be calibrated while mounted in the host system or it may be

removed and calibrated outside the system. The advantage of performing the calibration

while mounted in the host system is the user does not need to remove the SeaTRAX from

the system, which can be burdensome, and a simultaneous Mag-and-Accel Calibration

may be appropriate. The advantage of performing the calibration outside of the system is

it may be much simpler to obtain the desired range of pitch and roll.

Note: While the SeaTRAX is shown removed from the host system, the Accelerometer

Calibration may be performed with the SeaTRAX mounted in the host system.

Figure 5-2 shows the two basic starting positions for the recommended 18-point

calibration pattern. Starting with the SeaTRAX as shown on the left in Note: While the

SeaTRAX is shown removed from the host system, the Accelerometer Calibration may

be performed with the SeaTRAX mounted in the host system.

Figure 5-2, rotate the device about its z axis such that it sits on each of its 4 edges, taking

one calibration sample on each edge. Then place the SeaTRAX flat on the surface and

take a calibration sample, then flip it over (roll it 180°) and take another sample. Next,

starting with the SeaTRAX as shown on the right, take a calibration point with it being

vertical (0°). Now tilt the SeaTRAX back 45° and take another calibration point (+45°),

then tilt the device forward 45° and take another calibration point (-45°). Repeat this 3-

point calibration process for the SeaTRAX with it resting on each of its 4 corners. Note

that it is possible to perform an Accelerometer Calibration with as few as 12 sample