Now enter the zero offset sensitivity of the sensor by using the
and
button to select a digit and the
and
buttons to change the digit. When the value has been set, press
.If you do not know this, simply enter
all zeros.
You will be prompted with
dISP LO
. Press
to enter the display value required for the low input figure entered.
The value can be entered by using the
and
button to select a digit and the
and
buttons to
change the digit. When the value has been set press
.
You will be prompted with
InPut HI
, press.
Now enter the sensitivity figure supplied for the sensor by using the
and
button to select a digit and the
and
buttons to change the digit. When the value has been set press
.
You will then be prompted with
dISP HI.
Press
to enter the display value required for the high input figure
entered. The value can be entered by using the
and
button to select a digit and the
and
buttons to change the digit. When the value has been set press
.
You should now see
donE
displayed. This means the calibration was successful, press
to the IHH200 to normal
operation mode, with the new calibration data stored. If you see
FaiLEd
, then you will need to repeat the
calibration, checking that you have completed the procedure in the correct order, and that the sensor is connected
correctly.