Now enter the zero offset sensitivity of the sensor by using the

and

button to select a digit and the

and

buttons to change the digit. When the value has been set, press

.If you do not know this, simply enter

all zeros.

You will be prompted with

dISP LO

. Press

to enter the display value required for the low input figure entered.

The value can be entered by using the

and

button to select a digit and the

and

buttons to

change the digit. When the value has been set press

.

You will be prompted with

InPut HI

, press.

Now enter the sensitivity figure supplied for the sensor by using the

and

button to select a digit and the

and

buttons to change the digit. When the value has been set press

.

You will then be prompted with

dISP HI.

Press

to enter the display value required for the high input figure

entered. The value can be entered by using the

and

button to select a digit and the

and

buttons to change the digit. When the value has been set press

.

You should now see

donE

displayed. This means the calibration was successful, press

to the IHH200 to normal

operation mode, with the new calibration data stored. If you see

FaiLEd

, then you will need to repeat the

calibration, checking that you have completed the procedure in the correct order, and that the sensor is connected

correctly.