Instructions for configuration, method 2, To be completed if there is no scale factor, Untitled – FUTEK IPM500 (D500) Digital Display User Manual
Page 6
Revision 04
Instructions for Configuration, Method 2:
This method requires that you calculate the required scale and offset values and enter them into the
meter.
1. Apply Power. Press the Menu button and the display will read InPut. Press the Peak button to input
the signal type. Press the Peak button until the display reads Strn. Next, hit the Reset button, 20.0
should appear. If not, continue to press the Reset button until 20.0 appears. Press the Menu button to
store the input. SEtuP should appear on the display.
2. Press the Peak button, four numbers should appear '00_05'. Enter this setting into the display (if not
already showing) by selecting placement with the Peak button and numerical value with the Reset button.
Press the Menu button to save the meter settings.
Note: For older displays, enter the following configuration: 00005.
3. ConFG will appear. Begin to configure the meter by pressing the Peak button. Enter '020_0 using the
Peak
button to scroll through the digits (the digit you are on will flash) and the Reset button to increment
the flashing digit. Press the Menu button to save the meter configuration. FiLtr will appear.
Note: For older displays, enter the following configuration: 00000.
4. Press the Peak button in order to input your filtering options. Futek's default filter setting is '11100'.
Enter the numbers by pressing Peak to select the position and Reset to select the numerical value. Press
the Menu button to store filtering selections. dEC.Pt will appear.
5. dEC.Pt will allow the user to select the desired decimal point position. Press the Peak button and
'dddd.d' will appear. The Reset button will move the decimal point, once the decimal point position is
selected press the Menu button to save the position. SCALE will appear. If it does not appear ,continue
pressing the Menu button until it does.
6. Press the Peak button in order to enter the desired scale. The scale factor can be found on the label of
the sensor or the certificate provided with the sensor. Select the numeral position by pressing the Peak
button and the numerical value by pressing the Reset button. Press the Menu button to store the desired
scale. OFFSt will appear. If there is no scale factor on the instrument, set the scale to 1. Continue with
directions and also complete optional step 10 (below).
7. When OFFSt appears, press Peak. The default offset should be '00000', if an offset is desired enter it by
selecting the numerical position with the Peak button and the numerical value with the Reset button.
Press the Menu button to store the offset. Loc 1 will appear.
8. Press the Peak button and '00000' should appear. If not, change the values to '00000' by using the Peak
and Reset buttons. Press the Menu button to store Loc 1. Loc 2 will then appear. Repeat this step for Loc 2
and Loc 3. If locks are desired, refer to the 'Menu Lock' section for position of desired locks.
9. Press the Menu button until rESEt appears. The meter is now ready to use. The Tare button on the
back of the meter has been configured. Simply press the Tare button in order to zero your meter at any time
during use. This feature can be canceled/changed by changing the setup (refer to Owners Manual).
*It is possible to change any of the settings by pressing the Menu button and selecting the menu option by
pressing the Peak button and varying the settings (described above). It is necessary to continue to press the
Menu
button until rESEt appears.
To be completed if there is no scale factor
10. To use the sensor and make sure that the scale factor is 1, place a known load on the sensor (at least
75% of max load). The display will read out a number in mV. Take your initial load and divide it by the number
displayed. The load divided by the output number will become your scale factor. For every decimal place
chosen (step 5), multiply your scale factor by 10. Once the scale factor is calculated, press the Menu button
until SCALE appears on the display. Then press Peak and change the scale factor from 1 to the calculated value.
Hit the Menu button in order to store the new scale factor. The meter is now ready for use.
5