beautypg.com

FUTEK IPM500 (D500) Digital Display User Manual

Page 10

background image

Revision 04

3 – AL1 active high

AL2 active low

4 – AL1 active low

AL2 active low

5 – AL1 disabled

AL2 active low

6 – AL1 Active high

AL2 disabled

7 – AL1 active low

AL2 disabled

8 – AL1 disabled

AL2 disabled

00000 – Selects the Hysteresis mode or Band Deviation mode of alarms

0 – AL1 BAND Deviation

AL2 Band Deviation

1 – AL1 Hysteresis

AL2 Band Deviation

2 – AL1 Band Deviation

AL2 Hysteresis

3 – AL1 Hysteresis

AL2 - Hysteresis

4 – No deviation or hysteresis in menu

00000 – Number of reading taken in the alarm zone before the alarm goes off

0 – 1 reading

1 – 2 readings

2 – 4 readings

3 – 8 readings

4 – 16 readings

5 – 32 readings

6 – 64 readings

7 – 128 readings

4. ‘dEU1H’ will appear on the display, enter the setpoint at which you would like ‘alarm 1’ to be

activated. Press the MENU button.

5. ‘dEU2b’ will appear on the display, enter the setpoint at which you would like ‘alarm 2’ to be

activated. Press the MENU button.
Your alarm functions are now ready to use.