Pin menu, Login screen, Maintenance – Badger Meter M-Series M1000 User Manual
Page 30: Cleaning the flow tube and electrode, Troubleshooting

TROUBLESHOOTING
The M1000 mag meter is designed for many years of optimal performance . However, should it malfunction, there are certain
things that we recommend you check before contacting our Technical Support department or your local Badger Meter
Representative .
Errors & Warnings
OTEE:
N
The M1000 display flashes whenever an error is detected .
Description
Possible Cause
Recommended Action
Coil disconnected • Meter not connected .
• Connection to meter interrupted .
Check if meter is connected and make sure that
cable connection is not interrupted or contact
Badger Meter Technical Support .
Coil shorted
Coil cables shorted .
Check coil cables .
Empty pipe
Pipe may not be full .
Make sure that pipe is always filled at the
measuring point .
Medium with low conductivity .
Eventually calibrate new, see calibration of fluid
monitoring .
Cable broken or disconnected .
Check the cable for the empty pipe signal .
Range
Actual flow rate is exceeding the programmed full
scale by more than 100%
Reduce flow rate or increase the programmed full
scale .
Pulse output
Pulse rate exceed the maximum
Reduce pulse scale (pulse/unit) and/or reduce
pulse width configuration .
AD error
Input signal from detector too high .
Check the grounding scheme of the meter
installation . See
Excitation
frequency
The excitation frequency is too high for this
detector .
Decrease the excitation frequency in the Meter
Setup .
EEPROM
Configuration file is missing .
Contact Badger Meter Technical Support .
Configuration
Configuration file is corrupted .
Contact Badger Meter Technical Support .
Measure Timeout Measurement was not completed within specific
time .
Contact Badger Meter Technical Support .
Some frequently occurring situations:
Description
Possible Cause
Recommended Action
Meter does not
function
No auxiliary power .
Apply auxiliary power .
Fuse is defective .
Replace fuse .
Fluid is flowing,
however display
shows zero
Signal cable is not connected or connection is
interrupted .
Check signal cable .
Detector installed opposite to forward flow
direction (see arrow on type plate) .
Turn detector by 180° .
Connection cable for coils or electrodes mixed-up . Check connection cable .
Inaccurate
measurement
Wrong parameters .
Check parameters (detector, amplifier and size) per
Pipe not completely full .
Check if the measuring pipe is completely full .