Cover removal/reinstallation, Programming procedures, List item selection procedure – Badger Meter MR Transmitter User Manual
Page 13
MENU
Enters and exits programming mode . Press MENU once to change to programming mode . The mode
indicator on the display will change from RUN to PROGRAM .
Press
to scroll through the configuration choices in a bottom-to-top order . For numeric setup, this
button increments numeric values .
Use
to scroll through the configuration choices in a top-to-bottom order . For numeric setup, this
button moves the active digit to the right .
ENTER
Used to enter menus, to change configurations and to save programming information .
OTEE:
N
If any input value exceeds the meter’s capabilities, the LIMIT indicator will begin to flash indicating an invalid entry .
Press ENTER once to return to the entry screen to reenter the value .
Cover Removal/Reinstallation
It is necessary to remove the MR flow transmitter cover to access the programming keys . Use a Phillips screwdriver to remove
the four screws that hold the cover in place, turning them counterclockwise . When programming is completed, reinstall the
cover . To properly seat the built-in cover gasket, tighten the cover screws clockwise in a crisscross pattern as shown in
.
Flow Transmitter
3500 PSI/241 BARS MAX
OIL
1
2
3
4
Figure 8: Cover screw tightening sequence
Programming Procedures
The MR flow transmitter has been programmed at the factory according to the specifications that were provided at the time
of order . No further programming is required unless a change has occurred in the original specifications .
If programming is required, the MR flow transmitter allows two basic sets of programming procedures:
1 . List Item Selection
2 . Numeric Value Entry
List Item Selection Procedure
OTEE:
N
If you are already in PROGRAM mode and the selection to be viewed or changed is already displayed, proceed to step
3 below . If you are in PROGRAM mode and the selection to be viewed or changed is not displayed, press or
and repeat pressing until the desired selection appears . Proceed to step 3 .
1 . Press MENU . PROGRAM appears in the lower left-hand corner and DISPLAY appears .
2 . Press or
to move to the desired selection .
3 . Press ENTER to view the current selection .
4 . If the current selection is desired, press ENTER to confirm . The unit will automatically advance .
5 . If current selection must change, press or
to scroll through the available choices . Press ENTER to confirm your
selection . The unit will automatically advance .
6 . To exit programming, press MENU . The display will change to RUN mode .
User Manual