Er12 - CAL2 Voltage Error

1. Check that 50 mVDC is properly connected to the
instrument and is within the tolerance limits as
indicated in the CAL2 procedure of the Calibration
section (page 35).
2. Turn off the power to the instrument and open the
cover. Loosen the platen hold down screw and swing
the platen open. Inspect the Processor board
(Appendix A-1, page 50) to insure that the input
conditioning jumper JU4 is in the non-volt position.
3. Perform the CAL2 procedure as described in the
Calibration section (page 35).

Er13 - RTD CAL5 Input

1. Check that resistance device is of the correct value

Error

and properly connected to the instrument and is
within the tolerance limits as indicated in the CAL5
procedure of the Calibration section (page 37).
2. Turn off the power to the instrument and open the
cover. Loosen the platen hold down screw and swing
the platen open. Inspect the Processor board
(Appendix A-1, page 50) to insure that the input
conditioning jumpers are in the correct position for the
RTD input for the Terminal Board being calibrated.
For TB 4, JU4 should be in the non-volt position and
JU6 in the RTD position. For TB 5, JU5 should be in
the non-volt position and JU7 in the RTD position.
3. Perform the CAL5 procedure as described in the
Calibration section (page 37).

Er14 - Cold Junction

1. Perform the CAL3 procedure as described in the

Compensation Error

Calibration section (page 36).

Er15 - Ground Reference

1. Perform the CAL2 procedure as described in the

Tolerance Error

Calibration section (page 35).

Er16 - Program/ASEt Mode

1. Record all Program and ASEt mode parameters.

Checksum Error

Perform CAL 1 procedure as described in the
Calibration section (page 35). Re-enter the
Program and ASEt mode parameters.

Er17 - Calibration

1. Perform the calibration procedures that are

Checksum Error

needed for the input sensors that will be used (page
35).

Er21 - Pen Feedback

1. Perform the steps listed for pens not moving.

Error

Er36 - Incorrect Crystal

1. Turn off the power to the instrument, wait 5 seconds,

For Digital Comm.

then turn the power on.

Er37 - Incorrect Micro.

1. Turn off the power to the instrument wait 5 seconds,

For Digital Comm.

then turn the power on.

Er38 - Incorrect RAM

1. Turn off the power to the instrument, wait 5 seconds,

for profiler

then turn the power on.
2. Re-configure for non-profiler.