Troubleshooting – Thermo Fisher Scientific Alpha DO 500 User Manual

Page 43

background image

Instruction Manual

Alpha DO 500

37

8. Troubleshooting

Problem

Cause

Solution

Power on, but no
display

a) Loose connections
b) Power cables not in

correct polarity (+ & -)

c) Incorrect output voltage

of the power supply

a) Ensure cables make good

contact

b) Re-wire loop cables with

correct polarity

c) Use a power supply with

specified output voltage

Unstable dissolved
oxygen reading

a) Air bubbles in electrode
b) Dirty electrode

a) Tap electrode to remove air

bubbles

b) Clean electrode and

recalibrate

Oscillating
temperature readings

a) Electrical noise

interference


a) Ensure shield wire is

properly connected to pin 7

Slow response

a) Dirty / Oily electrode

a) Clean electrode

Blinking ATC

a) No temperature probe

connection during ATC
mode

a) Ensure temperature

sensing cable makes good
contact

Blinking electrode

annunciator

a) Error in calibration

a) Ensure calibration standard

solution is not
contaminated. Ensure
electrode is clean.

Or (Dissolved
oxygen)

a) Dissolved oxygen

electrode is not
connected

a) Ensure electrode makes

good contact with
Transmitter

Or (Temperature)

a) Temperature probe is

not connected when
ATC enabled

a) Ensure electrode makes

good contact with
Transmitter