5.8 Viewing Calibration Point

Calibration setting lets you view the high-point at which the calibration has been
performed in the Transmitter.

ENT

SETUP

HOLD

1

ENT

SETUP

HOLD

2

%

1

1

From measurement mode press ENT key to enter setup mode as

described in section 5.1. The LCD shows the first screen of setup mode
(COF). Press Ÿ or ź key to select viewing calibration settings screen
(CAL).

Press ENT key to view calibration settings (CAL).

2

The buffer annunciator

appears in the LCD. The upper display shows

the value of the high-point at which the calibration has been carried out.
The corresponding unit of measurement of calibration is displayed.

NOTE: If the Transmitter has not been calibrated yet for the currently

selected unit of measurement, the LCD shows an empty
reading (“---“ ) in the upper display.

Press ENT key to exit calibration settings screen. The Transmitter
reverts to SET CAL screen.

Press Ÿ or ź key to access other setup screens or press Ÿ and ź key
simultaneously
(escape) to return to measurement mode.