Instruction Manual

Alpha DO 500

6

2.3.2 Connecting Dissolved Oxygen Electrode

Refer the instruction sheet that comes with your Dissolve Oxygen electrode
to identify polarity (+ve and –ve) of DO wires.

1. Connect DO +ve wire to Pin 1 of SL1 connector

2. Connect DO -ve wire to Pin 2 of SL1 connector

2.3.3 Connecting Temperature Probe

For Automatic Temperature Compensated (ATC) readings, a 100

: Pt RTD

temperature probe (2-wire or 3-wire) can be connected to the Transmitter.

3-Wire Probe:

1. Connect PT100 compensate wire to Pin 5 of SL1 connector

2. Connect PT100 sense wire to Pin 6 of SL1 connector

3. Connect PT100 GND wire to Pin 7 of SL1 connector

2-Wire Probe:

1. Short Pin 5 & 6 of SL1 connector using a small piece of wire.

2. Connect PT100 sense wire to Pin 6 of SL1 connector

3. Connect PT100 GND wire to Pin 7 of SL1 connector