Instruction Manual

Alpha DO 500

31

5.9 Viewing Electrode Condition

The electrode condition (correction factor=K) is an indication of its current working
condition with respect to its original condition. The correction factor is calculated and
stored in the Transmitter each time calibration is done.

Electrode Correction Factor (K) = Theoretical reading of the electrode

Default

reading

of

the

electrode

ENT

SETUP

HOLD

1

ENT

SETUP

HOLD

2

K=

1

1

From measurement mode press ENT key to enter setup mode as

described in section 5.1. The LCD shows the first screen of setup mode
(COF). Press Ÿ or ź key to select viewing electrode condition screen
(ELE).

Press ENT key to view electrode condition (ELE).

2

The electrode annunciator appears in the LCD. The upper display

shows the current correction factor of the electrode (K)

NOTE: If the Transmitter has not been calibrated yet for the currently

selected unit of measurement, the LCD shows an empty
reading (“---“ ) in the upper display.

Press ENT key to exit the electrode condition screen. The Transmitter
reverts to ELE screen.

Press Ÿ or ź key to access other setup screens or press Ÿ and ź key
simultaneously
(escape) to return to measurement mode.