beautypg.com

7 out of range settings – Thermo Fisher Scientific Alpha DO 500 User Manual

Page 35

background image

Instruction Manual

Alpha DO 500

29

5.7 Out of Range Settings

When the Transmitter measures a dissolved oxygen value which does not fall within
the DO range (Zoom window) defined in the ‘Range Settings’, the reading is said to
be ‘out of range’. Out of range setting allows you to define the output current of the
Transmitter when the Transmitter measures an ‘out of range’ DO reading.

ENT

SETUP

HOLD

1

ENT

SETUP

HOLD

2

SETUP

HOLD

2

ENT

1

1

From measurement mode press ENT key to enter setup mode as

described in section 5.1. The LCD shows the first screen of setup mode
(COF). Press Ÿ or ź key to select out of range settings screen (SEt
org
).

Press ENT key to access out of range settings (SEt org).

2

The upper display shows ‘org’ and the lower display shows the last

configured out of range setting (‘On’ or ‘OFF’). Press Ÿ or ź key to
select the required choice.

- Select On to set output current to 3.8mA

- Select OFF to set output current to the boundary values:

That is: To set output current to 4mA when DO reading goes

bellow lower limit of the zoom window and,

To set output current to 20mA when the DO reading goes
above the upper limit of the zoom window.

Press ENT key to confirm your selection. The Transmitter reverts to SET
org
screen.

Press Ÿ or ź key to access other setup screens or press Ÿ and ź key
simultaneously
(escape) to return to measurement mode.