beautypg.com

3 calibration in% saturation of oxygen – Thermo Fisher Scientific Alpha DO 500 User Manual

Page 23

background image

Instruction Manual

Alpha DO 500

17

4.3 Calibration in% Saturation of Oxygen

The DO Transmitter is capable of calibrating up to 2 points for dissolved
oxygen. All new calibration values will automatically override existing
calibration data. Make sure the unit of measurement is set to %. (Refer
Section 5.2 for switching unit of measurements).

If you wish to abort the operation during DO calibration, press Ÿ and ź keys
simultaneously and the DO Transmitter reverts to measurement mode.

4.3.1 1-Point

Calibration

The 1-point calibration is done in air (100% oxygen saturation).

Measured value should be

within +/- 50 % of high-point

Calibration error occurs if measured value

is not within +/- 50% of high-point

during 1-point calibration

2

CAL

HOLD

1

ENT

ENT

ENT

CAL

HOLD

2

CAL

HOLD

2

CAL

HOLD

3

%

MEAS

°C

ATC

READY

4

CAL

HOLD

READY

ERR

%

%

+

1

1

From measurement mode press CAL key to enter calibration mode as

described in section 4.1. The LCD shows ‘CAL Do’. Press ENT key to
begin first calibration point.

2

Select 1-point calibration: The display shows ‘1Pt CAL’. If ‘2Pt CAL’ is

shown, press Ÿ or ź key to select ‘1Pt CAL’.

Press ENT key to confirm the selection.

3

The buffer annunciator

appears in LCD. The lower display shows

100.0’. Allow the electrode to equilibrate in air. The upper display shows
the measured value of dissolved oxygen saturation. Allow the reading to
stabilize. LCD shows ‘READY’ annunciator when the reading is stable.

Press ENT key to confirm the reading.

NOTE: The transmitter does not accept the calibration and shows

calibration error if the displayed (measured) value is not within
50% to 150%.

4

The calibration is completed. The DO Transmitter reverts to

measurement mode.