16 ADDENDUM 4: METER FACTORY DEFAULT SETTINGS
Type Parameter
Default
Remarks
P1.1
Viewing calibration solution
––
No calibration data for first range
P1.2
––
No calibration data for second range
P1.3
––
No calibration data for third range
P1.4
––
No calibration data for fourth range
P1.5
––
No calibration data for fifth range
P2.1
Viewing probe data
1.000
No offset for effective cell constant (first range)
P2.2
1.000
No offset for effective cell constant (second range)
P2.3
1.000
No offset for effective cell constant (third range)
P2.4
1.000
No offset for effective cell constant (fourth range)
P2.5
1.000
No offset for effective cell constant (fifth range)
P3.1
Ready indicator / Auto-hold READY/ON
Ready
indicator on; Auto-Hold off
P3.2
Select °C/°F
°C
Degrees Celsius
P3.3
ATC on or off
ATC YES
––
P3.4
TDS factor
0.5
Adjustable from 0.4 to 1.0
P4.1
Temperature coefficient
2.1 % per ° C
Adjustable from 0 to 10%
P4.2
Normalization temperature
25° C
Adjustable from 15 to 30° C
P5.0
Auto or manual calibration
Auto YES
Auto or manual
P5.1
Single or multi-point
calibration
Single-point YES
Single or Multi-point calibration
P6.0 Cell
constant
1.0
Select from k=0.1, 1.0 or 10
P7.0
Factory default
No
Retains your current settings