Instruction Manual

CON 510

7.6

P5.0 Mode of calibration

.

SETUP

SETUP

SETUP

SETUP

SETUP

ENTER

RANGE

ENTER

RANGE

ENTER

RANGE

CAL

MEAS

The CON 510 bench meter allows you to perform automatic (only for Conductivity
mode) or manual calibration (applicable to both Conductivity/TDS mode). Similarly you
may choose to perform either single-point or multi-point calibration (up to 4 preset
calibration standards in automatic calibration mode). Refer to the Section 4 for details
on calibration.

1. From measurement mode, press SETUP to enter into setup [SETUP] mode.
2. Use

MI/▲ or MR/▼ key to scroll through main-menu “ACAL P5.0”.

3. Press ENTER key to enter into this main-menu for selection of Automatic

Calibration either “nO” (Manual) or “YES” (Automatic) using MI/▲ or MR/▼ key.

4. Once the option is selected, press ENTER key to confirm. The next display shows

the option of Single-point Calibration (SPC) or Multi-point Calibration.

5. Use

MI/▲ or MR/▼ key to select “YES” if you wish to perform Single-point

Calibration and “nO” to perform Multi-point Calibration.

6. Press

ENTER key to confirm after appropriate selection is made and to return to

the SETUP main-menu P5.0. Press CAL/MEAS key to return to measurement
[MEAS] mode.