6 p5.0 mode of calibration – Thermo Fisher Scientific CyberScan CON 510 User Manual
Page 24

Instruction Manual
CON 510
7.6
P5.0 Mode of calibration
.
SETUP
SETUP
SETUP
SETUP
SETUP
ENTER
RANGE
ENTER
RANGE
ENTER
RANGE
CAL
MEAS
The CON 510 bench meter allows you to perform automatic (only for Conductivity
mode) or manual calibration (applicable to both Conductivity/TDS mode). Similarly you
may choose to perform either single-point or multi-point calibration (up to 4 preset
calibration standards in automatic calibration mode). Refer to the Section 4 for details
on calibration.
1. From measurement mode, press SETUP to enter into setup [SETUP] mode.
2. Use
MI/▲ or MR/▼ key to scroll through main-menu “ACAL P5.0”.
3. Press ENTER key to enter into this main-menu for selection of Automatic
Calibration either “nO” (Manual) or “YES” (Automatic) using MI/▲ or MR/▼ key.
4. Once the option is selected, press ENTER key to confirm. The next display shows
the option of Single-point Calibration (SPC) or Multi-point Calibration.
5. Use
MI/▲ or MR/▼ key to select “YES” if you wish to perform Single-point
Calibration and “nO” to perform Multi-point Calibration.
6. Press
ENTER key to confirm after appropriate selection is made and to return to
the SETUP main-menu P5.0. Press CAL/MEAS key to return to measurement
[MEAS] mode.
20