P4.2 Adjusting the Normalization Temperature

SETUP

o

C

SETUP

o

C

..

.

o

F

.

..

.

o

C

o

F

ENTER

RANGE

ENTER

RANGE

CAL

MEAS

Your meter will normalize its conductivity measurements to a standard temperature that
you can select. You can adjust the normalization temperature from 15 to 30

°C (59 to

86

°F). Meter default is 25.0 °C (77 °F).


1. From measurement mode, press SETUP to enter into setup [SETUP] mode.
2. Use

MI/▲ or MR/▼ key to scroll through the SETUP main-menu P4.0 and press

ENTER thrice repeatedly until you can view the sub-menu P4.2 which the upper
display shows “t.nr” on and “P4.2” on the lower display.

3. Press

ENTER key to make change to the displayed value. The upper display

shows the normalization temperature and the lower display shows “t.nr”.

4. Press

MI/▲ or MR/▼ key to set the normalization temperature (in the specific unit

of temperature which is set at SETUP sub-menu P3.2).

5. Press

ENTER to confirm selection and to return to the SETUP main-menu P4.0.

Press CAL/MEAS key to return to measurement [MEAS] mode.