beautypg.com

7 manual calibration (conductivity/tds), 1 conductivity manual calibration, 2 tds manual calibration – Thermo Fisher Scientific CyberScan CON 510 User Manual

Page 12: 7 manual calibration

background image

Instruction Manual

CON 510

4.7

Manual Calibration (Conductivity/TDS)

CAL

MEAS

o

C

ATC

READY

.

uS

uS

CAL

uS

o

C

ATC

MEAS

.

uS

READY

MODE

ENTER

RANGE

CAL

MEAS

CAL

MEAS

The meter accepts either single point or multi-point (up to 5 points) manual calibration
with maximum of 1 point per measurement range. Freshly prepare your standards
solution before calibration. Refer to the table in Section 4.1 for more details of the
recommended standards calibration range.

You can offset or manually adjust the Conductivity or TDS reading up to ±40% from its
default setting. If your measured value differs by more than ±40%, clean or replace probe
as needed. Refer to Section 8 for Probe Maintenance.

NOTE: You need to set your desired option i.e. number of calibration points in the
SETUP main-menu P5.0 before performing calibration.


4.7.1 Conductivity

Manual

Calibration

1. If necessary, press the MODE key to select conductivity mode.
2. Rinse the probe thoroughly with de-ionized water or a rinse solution, and rinse with a

small amount of calibration standard.

3. Dip the probe into the calibration standard. Immerse the probe tip beyond the upper

steel band. Stir the probe gently to create a homogeneous sample. Allow time for the
reading to stabilize.

4. Press

CAL/MEAS key to enter conductivity calibration mode. The [CAL] indicator

will appear in the upper right corner of the display.

5. Use

MI/▲ or MR/▼ key to adjust the value on the upper display to match the value

of the calibration standards.

6. Press

ENTER to confirm calibration value. The meter returns to the [MEAS]

measurement mode.

7. To abort calibration without confirming, press CAL/MEAS key to revert back to

measurement mode. No calibration is performed at this stage.

9. For single-point calibration the meter automatically reverts back to the measurement

mode.

10. For multi-point calibration repeat steps 2 to 6 for the second to fifth calibration value

in specific measuring ranges. To abort calibration at any calibration points, press
CAL/MEAS once and the meter reverts to measurement mode.



4.7.2

TDS Manual Calibration

The calibration sequence is exactly the same as Conductivity manual calibration. Ensure that you set the correct TDS
factor before calibration. Refer to Section 7.4 for details.















8