Thermo Fisher Scientific CyberScan CON 510 User Manual
Page 21
Instruction Manual
CON 510
P3.2 Selecting
°C or °F
SETUP
o
C
.
SETUP
o
F
.
CAL
MEAS
ENTER
RANGE
ENTER
RANGE
You can select between
°C and °F as unit of measure for temperature readings. Meter
default is
°C.
1. From measurement mode, press SETUP key to enter into Setup [SETUP] mode.
2. Use
MI/▲ or MR/▼ key to scroll through the SETUP main-menu P3.0 and press
ENTER until you can view sub-menu P3.2 on the display.
3. Use
MI/▲ or MR/▼ key to toggle between
°C and °F.
4. Press
ENTER to confirm selection and you will automatically return to the SETUP
main-menu P3.0. Press CAL/MEAS key if you wish to return to measurement
[MEAS] mode.
NOTE: Similarly if you are accessing SETUP main-menu P3.0, pressing ENTER twice
will take you to this SETUP sub-menu P3.2.
SETUP
SETUP
SETUP
o
C
ATC
o
C
ATC
.
o
C
ENTER
RANGE
ENTER
RANGE
ENTER
RANGE
CAL
MEAS
P3.3 Selecting Automatic or Manual Temperature Compensation
This Program P5.3 allows you select between Automatic Temperature Compensation
(ATC) and Manual Temperature Compensation. Meter default is ATC.
1. From measurement mode, press SETUP key to enter into setup [SETUP] mode.
2. Use
MI/▲ or MR/▼ key to scroll through the main-menus “COF P3.0” and press
ENTER key repeatedly until you come to the SETUP sub-menu P3.3 which the
upper display shows “ATC” and “P3.3” on the lower display.
3. Press
ENTER key again. The upper display shows “ATC” and the lower display
shows “YES” or “NO”.
4. Use
MI/▲ or MR/▼ key to select the Automatic Temperature Compensation on or
off. YES = ATC on; NO = ATC off
5. Press
ENTER key to confirm selection and to return to the SETUP sub-menu P3.3.
Press CAL/MEAS key to return to measurement [MEAS] mode.
17