beautypg.com

Addendum 3: calculating temperature coefficients – Thermo Fisher Scientific CyberScan CON 510 User Manual

Page 32

background image

Instruction Manual

CON 510

15 ADDENDUM 3: CALCULATING TEMPERATURE COEFFICIENTS

To determine the temperature coefficient of your sample solution use this formula:

Where:

tc = Temperature coefficient

25 = 25 °C

C

T1

= Conductivity at Temp 1

C

T2

= Conductivity at Temp 2

T

1

= Temp 1

T

2

= Temp 2

NOTE: A controlled temperature water bath is ideal for this procedure.

1. Immerse the probe into a sample solution and adjust the temperature coefficient to 0% (that is, no compensation) by

performing the following:
A. From measurement mode, press the SETUP key to enter into [SETUP] mode.
B. Use

MI/▲ or MR/▼ key until the lower display shows “P5.0”.

C. Press

ENTER key twice. The lower display reads tCO and the upper display shows the temperature coefficient

value.

D. Press

the

MR/▼ key until the upper display shows 0.0.

E. Press

ENTER key to confirm the value.

F. Press

CA/MEAS key to return to measurement mode.

2. Wait for 5 minutes. Note T

1

and C

T1

(conductivity at T

1

).

3. Condition the sample solution and probe to a temperature (T

2

) that is about 5 °C to 10 °C different from T

1

, and note

the conductivity reading C

T2

.


NOTE:
Record your results for future reference. Ideally T

1

and T

2

should bracket your measurement temperature, and

should not different by more than 5 °C.


4. Calculate the temperature coefficient of your solution according to the formula shown above.
5. Enter the temperature coefficient you calculated into the meter by repeat step A to C. Use MI/▲ or MR/▼ key to set

the desired value. Press ENTER key to confirm selection and then press CAL/MEAS key to return to measurement
mode.

The calculated temperature coefficient will not be applied to all the meter readings.

28