beautypg.com

8 p7.0: resetting to factory default settings – Thermo Fisher Scientific CyberScan CON 510 User Manual

Page 26

background image

Instruction Manual

CON 510

7.8

P7.0: Resetting to factory default settings

SETUP

SETUP

SETUP

.

ENTER

RANGE

ENTER

RANGE

CAL

MEAS

Program 7.0 allows you to reset all parameters to factory default settings. This clears all
calibration data, memory, and any other setup functions you might have changed.
IMPORTANT: Once activated the meter’s settings and calibration data will be erased
and always exercise caution as meter reset is not reversible.

1. From measurement mode, press SETUP key to enter into setup [SETUP] mode.
2. Use

MI/▲ or MR/▼ key to scroll through the main-menu “rSt P7.0” and press

ENTER key to enter into the selection options for the meter reset.

3. Use

MI/▲ or MR/▼ key to toggle between NO and YES.

NO retains current settings; YES resets to factory default settings.

4. Press

ENTER key to confirm selection and to return to the measurement mode.

Otherwise press CAL/MEAS key to return to measurement [MEAS] mode without
resetting to factory default.


NOTE:
See Addendum 4 for a table of factory default settings.


22