2. Turbidity-free water can be obtained by filtering deionized water through a < 0.2μm
filter membrane with precision-sized pores.
3. Measure the volume of the sample before dilution and record the value in ml (Vs).
4. Take a known volume (Vd) of dilution water and add it to the sample.
5. Pour 10 ml of the diluted sample in a clean vial and measure the turbidity of the
diluted sample. Record this value in NTU (Td).
6. Calculate the true turbidity (T) of the original sample - in NTU - using the following
formula:
T = Td * (Vs + Vd) / Vs
Example:
Dilute 20 ml of the original sample (whose turbidity is above 1000NTU)
with 50 ml of dilution water.
Measure the turbidity of the diluted sample.
If the reading is 300 NTU, the turbidity of the original sample is 1050
NTU. (In this case: Td=300NTU, Vs=20ml, Vd=50ml, so T = 300 *
(20+50) / 20 = 300*70/20 = 21000/20 = 1050)
10. Warranty
The TN-100/ T-100 meter is supplied with a
2-year warranty from manufacturing
defects and calibration standards for
6 months.
If repair or adjustment is necessary and has not been the result of abuse or misuse
within the designated period, please return – freight pre-paid – and correction will be
made without charge. Eutech Instruments/ Oakton Instruments will determine if the
product problem is due to deviations or customer misuse.
Out of warranty products will be repaired on a charged basis.
Exclusions
The warranty on your instrument shall not apply to defects resulting from:
•
Improper or inadequate maintenance by customer
•
Unauthorized modification or misuse
•
Operation outside of the environment specifications of the products
Waterproof Seal: Opening the instrument enclosure (excluding the battery
compartment) may void the warranty.