Instruction Manual

TN-100/ T-100

8

2.4

Restoring Factory Calibration

The TN-100/ T-100 turbidimeter allows you to reset the meter back to the factory

default calibration values. This feature is extremely useful when there are errors in

calibration or when you have new calibration standards.
1. With the meter switched off, press and hold ▲ key.

2. Switch on the meter using

ON/OFF. Release ▲ key when “URST” (User -Reset)

appears in the display.

3. The display shows “URST” for about 2 seconds and then shows “No”.

4. Use ▲ or ▼ key to select between “YES” or “No”.

YES = To restore meter back to factory calibration values

No = To retain last calibrated values

5. Press

READ/ENTER key to confirm your selection. The meter performs the reset if

“YES” is selected.

6. The display flashes “--Rd—“about 10 times indicating that the meter goes to

measurement mode.

Figure 6 shows the sequence for restoring factory calibration values.