For Continuous Measurement:

1. Make sure the meter is sitting on a flat and level surface and is in measurement

mode. The display shows the last measured value or “STbY”

2. Place the sample vial in the sample well.

3. Press

READ/ENTER key and hold. See Figure 13

4. Wait for the reading to stabilize before rotating the sample vial.

NOTE: During continuous measurement, the display is updated every 2 seconds.

The displayed reading may not be the actual turbidity value. For accurate

measurement, use single-shot measurement.

You can rotate the sample vial for indexing purpose. See

Section 8.5 - Indexing a Vial

on page 20 for more information.

5. Once you release

READ/ENTER key, the meter automatically performs a single-

shot measurement.












NOTES:
1. After a measurement is completed and the display is updated, a 4-

second recovery time occurs before the meter can perform another

function. If any key is pressed during the recovery time, the meter will

perform the corresponding action at the end of the recovery period.

2. When performing a measurement, if the meter detects stray light

exceeding the amount equivalent to 0.02 NTU in the low range, the

measurement is immediately aborted and an error message [ERR 8] is

displayed. Ensure the vial sits properly into the sample well. Press the

ENTER/READ key to re-do the measurement.

Figure 13: Continuous Measurement

Reading flashes and

updates every 2 seconds

SAMPLE

Unknown

NTU

Place sample

vial into the

sample well

NTU

READ

ENTER

Measurement Mode

Press key and hold