Instruction Manual

TN-100/ T-100

12

Measurement Procedure

1. Place TN-100/ T-100 turbidimeter on a flat and level surface.

2. Place the sample vial inside the sample well and align the vial’s index mark with the

meter’s index mark. See

Figure 10.

Figure 10: Align sample vial’s index mark with meter’s index mark

3. Push the vial until it is fully snapped in.

4. Cover the vial with the light shield cap.

5. Turn on the meter by pressing the

ON/OFF key.

6. After the power-up sequence, the meter

goes to measurement mode and the

display blinks “--Rd--“ for about 10 times.

See Figure 11.

7. The measured reading appears in the

display.

8. If necessary, place the second sample vial

into the sample well. Remember to align

the vial’s mark with the meter’s index

mark.

9. Press

READ/ENTER key. The display

blinks “--Rd--“ for several times and

measured reading appears.

10. Repeat steps

Error! Reference source

not found. through 9 for all of your

samples.

Align index mark on

the sample vial with

the index mark on

the TN-100/ T-100

turbidimeter and

push till it is fully

snapped in

IMPORTANT:

Place meter on a flat

and level surface.

DO NOT hold it on

hands while operating

the meter. It may lead to

inaccurate readings

Light shield cap for

covering the vial during

measurements

Figure 11:

Reading Turbidity Value

“-Rd-” flashes about

10 times

SAMPLE

Unknown

NTU

Place sample

vial into the

sample well

NTU

READ

ENTER

Measurement Mode