beautypg.com

Addendum 2: guide to good measurement technique, 1 maintain sample vials in good condition, 2 match sample vials – Thermo Fisher Scientific TN100 Turbidimeter User Manual

Page 26: 3 degassing, Addendum 2, Guide to good measurement technique, Maintain sample vials in good condition, Match sample vials, Degassing

background image

Instruction Manual

TN-100/ T-100

21

9. Addendum 2:

Guide to Good Measurement Technique

Turbidity is a very complex analytical measurement which can be affected by many

factors. Some are inherent in the instrument’s design such as angle of detection, light

beam aperture, incident beam wavelength and color sensitivity of the photocell.
However, there are other factors such as stray light, air bubbles and care of vial, which

can be prevented through proper care of equipment and accessories, and in the

operating procedure for measurement. Here are some points you may want to note:

9.1

Maintain sample vials in good condition

Sample vials must be meticulously clean and free from significant scratches. It should

be treated on the outside with a thin coat of silicone oil. This is to mask minor

imperfections and scratches that may contribute to stray light. Sample vials should be

handled only by the top to avoid dirt accumulation (or deposits) and fingerprints that

might interfere with the light path.
More information is found in Section 5.1 - Vials – Handling, Cleaning and Care on page

16.

9.2

Match Sample Vials

Best accuracy and repeatability of turbidity measurement are achieved using a single,

indexed vial. However, for more convenience, different vials can be used for

measurement provided their readings with the same solution are matched. That is, the

meter gives identical readings or within the specified repeatability and accuracy of the

meter.
Select a few vials. After the sample vials are cleaned, fill them with ultra-low turbidity

water. Allow the sample vials to stand and for air bubbles to rise. Polish sample vials

with silicone oil and take turbidity measurement at several points while rotating it in the

sample well. Find the position where turbidity reading is the lowest and index it for each

vial. Whenever these sample vials are used, use the indexed mark to position each vial

into the sample well. Choose those vials that match the readings.

NOTE: Not all vials can be matched because of some manufacturing variations.

9.3

Degassing

Air or other trapped gases should be removed before measurement. Degassing is

recommended even if no bubbles are visible. There are three methods commonly used

for degassing:

Addition of a surfactant: This involves adding a surfactant to the water

samples to lower the surface tension of the water, thereby releasing trapped

gasses.

Application of a partial vacuum: Partial vacuum can be created by using

simple syringe or vacuum pump. (This is only recommended for ultra-low

turbidity measurement).