Instruction Manual

TN-100/ T-100

13

NOTES:
1. NEVER pour liquid directly into the sample well of the instrument. Always

use a vial. The instrument will only accurately measure the turbidity of a

sample when vials sealed with the black caps are used. The black cap

serves as both seal and a light shield.

2. NEVER attempt to clean the sample well. The optics may be damaged.
3. For battery conservation, the instrument automatically powers off 20

minutes after the last key pressed.

3.3

Single-Shot or Continuous Measurement

You can use TN-100/ T-100 turbidimeter to take a single reading or perform continuous

measurement. The latter is only used for indexing the vials. See

Section 8.5 - Indexing

a Vial on page 20 for more information.

For Single-shot Measurement:

1. Make sure the meter is sitting on a flat and

level surface and is in measurement

mode. The display shows the last

measured value or “STbY” after exiting

calibration mode.

2. Place sample vial in the sample well.

3. Cover the vial with the light shield cap.

4. Press

READ/ENTER key and release

immediately (<0.3 seconds). See Figure

12.

5. The display blinks “--Rd--“ for about 10

times and then display the measured

value.





Figure 12: Single-shot Measurement

“-Rd-” flashes about

10 times

SAMPLE

Unknown

NTU

Place sample

vial into the

sample well

NTU

READ

ENTER

Measurement Mode

Press key and

release immediately