Instruction Manual

TN-100/ T-100

11

Preparation of Sample Vial

1. Obtain a clean and dry sample vial

2. Take care to handle the sample vial by the top.

3. Rinse the vial with approximately 10 ml of the sample

water, capping the vial with the black screw cap and

gently inverting it several times. Discard the used

sample and repeat the rinsing procedure two more

times.

4. Fill the rinsed vial with the remaining portion

(approximately 10 ml) of the grab sample up to the

mark indicated in the vial. Cap the vial with the

supplied black screw cap.

5. Wipe the vial with the soft, lint-free cloth supplied.

Ensure that the outside of the vial is dry, clean and

free from smudges.

6. Apply a thin film of silicone oil (supplied) on the sample vial (see Figure 9).

7. Wipe with a soft cloth to obtain an even distribution over the entire vial’s surface.

NOTES:
1. The purpose of oiling the vial is to fill small scratches and to mask the

imperfection in the glass.

2. Do not apply large quantity of oil as this may collect dirt and dust

8. The sample vial is now ready to be inserted into the sample well of the meter for

measurement.

Figure 9: Apply a thin film of silicon oil



Figure 8: Sample Vial