Techne Prime Elite User Manual
Page 49
47
Note: If the user attempts to send a program to unit
while it is already running, the warning screen
shown opposite will be displayed.
• Touch OK to exit.
THE PRE-RUN SCREEN
The Pre-run Screen shows a graphical representation of the program allowing for a rapid check of the
temperature parameters.
The centre of the screen shows each temperature
step and hold time (not to scale). Stages are
indicated by a white bar spanning the steps within
the stage.
The bottom of the screen shows the program name,
the date and the expected end time if it was started
immediately.
• Touch Start to start the program.
• Touch Edit to edit the program.
• Touch Back to return to the previous screen.
Note: The temperature log will remain blank until
the program has been started.
If a step has a gradient programmed into it, the bar
will be shown thicker and indicate the upper and
lower temperatures of the gradient range.
If a step has an incremented or decremented
temperature, this will be indicated by a white spot
next to the temperature.
If a step has an incremented or decremented time,
this will be indicated by a white spot next to the
time.
Gradient
Inc/dec
temperature
Inc/dec time
Gradient
Inc/dec
temperature
Inc/dec time