31
INSTRUMENT SETTINGS
Temperature Log Recording
To record the temperature logs (block temperature
profile) of each run, touch the button to set this to
Always. If temperature logs are not required, set to
Never.
Auto Resume On Power Loss
This determines the time limit after which the unit
will restart after a power failure. For example, if the
Auto Resume is set to 30 minutes, the unit will
restart at the beginning of the step at which it was
interrupted if the power failure lasted for 30 minutes
or less but not if it was for longer.
Setting Auto Resume to Always will always allow
the unit to re-start irrespective of the length of the
power failure.
Date and Time Settings
Use these settings to set the current date, time and
date format on the unit.
• When set, touch Back to return to the Settings
menu.
Date
• Touch Set Date in order to enter the current
date.
• Use the + and – keys to set the date values.
• When the correct date has been set, touch OK
to return to the previous screen.
• Touch Cancel to return to the previous screen if
no changes are required.