beautypg.com

General software features – Techne Prime Elite User Manual

Page 22

background image

20

GENERAL SOFTWARE FEATURES

INTRODUCTION

The Prime Elite software allows the user to quickly and easily create thermal cycling programs using a

simple and clear touch screen format. The software is structured into five modules which are accessed

from the Home Screen. These allow the user to:

• Run an existing program, selecting from

templates or programs saved on the unit.

• Create a new program.

• Set instrument and program defaults and

preferences. Includes calculator tools.

• View the current status of connected units.

• View and search for files stored on the unit.

TEMPLATE PROGRAMS

To help with programming, a number of program templates are installed on the unit which are

available to copy and edit or can be run directly without changes. Details of the installed program

templates are given below.

Instrument and program defaults

Parameter

2 Step Template

3 Step Template

RT-PCR Template

Heated Lid

105ºC

105ºC

105ºC

Heated lid before program

On

On

On

Sample cooling

On

On

On

Pause before program

Off

Off

Off

Initial denaturation

94ºC, 05m00s

94ºC, 05m00s

94ºC, 05m00s

Hot start

Off

Off

Off

Final extension

72ºC, 05m00s

72ºC, 05m00s

72ºC, 07m00s

Final hold

10ºC

10ºC

10ºC

Thermal cycling conditions

2 Step Template

3 Step Template

RT-PCR Template

Stage 1
Number of cycles

30

30

1

Step 1

94.0ºC, 00m30s

94.0ºC, 00m30s

45.0ºC, 40m00s

Step 2

60.0ºC, 01m00s

55.0ºC, 00m30s

94.0ºC, 05m00s

Step 3

72.0ºC, 00m30s

Stage 2
Number of cycles

40

Step 1

94.0ºC, 01m00s

Step 2

55.0ºC, 00m50s

Step 3

72.0ºC, 01m00s