Select the program required, using the cursor
keys to view all the available programs.

Press "Run".

The run screen is then displayed; if the heated
lid is on then the screen will show the
message "Waiting Heated Lid" and the time
until the lid will be ready. It is possible to stop
or pause the program.

If "Stop" is pressed then it is necessary to
confirm that that the program is to be stopped
by pressing "Stop" again.

If pressing "Stop" was an error then pressing
"Continue" will enable the program to
continue from where it was stopped.

If "Pause" was selected then press "Unpause"
to restart the program. The pause function
might be required for a manual hot start.

Program

3 STEP PCR TEMPLATE

Last Modified

01/11/1998

00:00:00

Heated lid

105.0°C

Heated lid before program

On

Pause before program

Off

Initial denaturation 05m00s 94.0°C

Hot Start

Disabled

Number cycles 30

Exit

Print

Run

Edit

Delete

Copy

▼ ▼

Select Program

14526 bytes of memory free

b 2 STEP PCR TEMPLATE

b 3 STEP PCR TEMPLATE

b ICE BUCKET

b LITIGATION 15 DEGREES

b RT PCR TEMPLATE

New

Exit

▼ ▼

Program

B ACTIN

Waiting Heated Lid 03m40s

Pause

Stop