beautypg.com

Terminals & switches, Set-up – SHIMPO FGS-250PV User Manual

Page 5

background image

5

Terminals & Switches

NO.DESIGNATION

1. Emergency Stop

2. Plug receptacle for force gauge

3. Power Switch

4. Power Cable

5. OUT: Permits transfer of force data from FGV, DFS or DRI series of force gauges and distance data from

programmed set points

6. IN: Permits communication between digital gauges and the test stand

7. LENGTH: Permits connection for analog output voltage related to distance

Set-Up

Connect all cables and power cords before applying

power to the test stand and force gauge. Apply power

to the force gauge before applying power to the test

stand; this is necessary so that the test stand will

recognize which Shimpo force gauge is being used.

To identify whether the stand is in metric or English units,

turn the power on (power SW is located in the back of

the unit); metric units will display one decimal place (0.0),

while English units display two decimal places (0.00).

To switch from English to metric units, push “SPEED” and

“RST/!” switches while turning the power on. To return

to English units, turn the test stand off, then push

“SPEED” and “TIME/$” switches while turning the power

back on.