1.3 Key Functions

All keys are capacitive touch.

ON/OFF: Push for 2 seconds to power On or Off


During Measurement: Print the current force value or

store data, depending on the key setting. (See 2.5.7 for
key

setting)

In Menus: Back or quit.

During Measurement: Enter the menus.

In Menus: Select or Enter

During Measurement: Track mode, tares weight of

attachment. In Peak & Auto Peak modes, resets the

peak

value.

In Menus: Moves selection up or increases the value.

During Measurement: Changes the measure mode from

Track, Peak or Auto Peak

In Menus: Moves selection down or decreases the

value.

1.4 Modes

Track: Real time, live measuring mode.
Peak: Peak readings will not changed until a higher value is mea-
sured.
Auto Peak: When the Peak Time is up, resets the peak value au-
tomatically. (See 2.2.4 for Peak Time.)

2. ADVANCED MENU OPTIONS

2.1 Menu Structure

The FG-7000 Series Force Gauge has multi-level menu interface
(Table 2-1) that enables simple navigation and programming.

Table 2-1

From the home screen, touch “MENU” to enter the main menu.
(Figure 2-1)

Figure 2-1a

Figure 2-1b

2.2 Measurement

The Measurement menu contains six selectable items: Unit,
Group, Tolerance, Peak Time and Alarm. (Figure 2-3)

Figure 2-3a

Figure 2-3b

2.2.1 Unit

The measuring unit can be selected under this menu. Differ-
ent range models may have different unit selection capabili-
ties. Touch “ZERO” or “MENU” keys to shift to the next selec-
tion. Press “LOG” to cancel or touch “MENU” to confirm and
exit. (Figure 2-4)

Figure 2-4

2.2.2 Group
When several test samples need to be measured, the sam-
ples can be coded into groups. The range is 01-99. When set
to “00”, become, “01” automatically. Press “ZERO” to adjust
the value, touch “MODE” to shift to the next position. Touch
“LOG” to cancel; press “MENU” to confirm and exit. (Figure
2-5)

Figure 2-5

2.2.3 Tolerance
In the Tolerance menu, program high and low limit values to
enable ok/ov testing. The lower limit value cannot be greater
than the upper limit value, and neither limit value can be great-
er than 110% of the rated capacity. Press “ZERO” to adjust the
value, touch “MODE” to shift to the next position. Press “LOG”
to cancel; touch “MENU” to confirm and exit. (Figure 2-6)

Figure

2-6

Menu

Measurement

Unit

Group

Tolerance

Test Mode

Peak Time

Alarm

Memory

Storage Mode

Browse All

Browse Selected

Delete Selected

Delete All

Printing

Print Recent

Print Selected

Print All

System

Display Mode

Power Off

Backlight

Key Tone

Date/Time

Password

Key Setting

Default Setting

Language

Calibration

Information

2