KROHNE BM 100A Handbook User Manual
Page 43

 
 
Handbook BM 100 A
43
If the level pulse is too high 2 solutions are possible: 
 
• Modify user function 1.5.3 Detection Delay ignore all signals near the flange 
Increase the configured value. This will force the gauge to ignore reflections in the distance set from 
the flange. Note that this will reduce the measurement range. 
 
• Modify user function 1.5.1 Level Threshold: to reduce interference along the probe length 
Increase the voltage amplitude manually in Fct. 1.5.1. The threshold descends in 2 metre steps to 
take into account loss of pulse strength over distance so the correct level behind the obstruction 
may be detected. See the procedure below: 
 
 
If measured level is too low compared its real level: 
 
The gauge may have found that the short circuit gives off the strongest pulse and assigns this as 
being the current level. Lowering the gain and /or the voltage amplitude manually in Fct. 1.5.1. may 
help but too low a threshold value may result in conflicting non-product reflections and unstable 
readings. In the case of very low dielectric products, contact KROHNE service personnel or refer to 
the BM100 A service manual for advice on converting to TBF measurement. 
 
 
 Procedure (example where the level measured is too low) 
• read off displayed peak amplitude 
• modify level threshold value
Step Action
Press buttons to finish 
step 
Information 
displayed at the 
end of each step 
1
Start from Operating Mode screen
n/a
 6750 
LEVEL mm 
2
From operating mode go to fct. 1.5.1 Level 
Threshold 
→,→,↑,↑,↑,↑,→
Fct.
1.5.1
LEVEL
3
Enter function ( current value: level) to first 
read off peak amplitude. The top value 
gives the peak amplitude in volts. 
→
 3.80 
 GAIN2 
4
Press enter to see (& modify) configured 
level threshold in volts and the minimum 
gain used 
↵
3.80
MIN G 2
5
Modify the threshold value so that value is 
no less than 0.5V below the peak amplitude 
→, ↑,↑,↑,↑,↑,↑,→,↑,↑,↑,↑
3.44
MIN G 2
6
Exit data set field, function, sub menu and 
menu. Store configuration and return to 
operating mode. Check that the gauge can 
measure in the required measuring range. 
↵,↵,↵,↵
 6750 
LEVEL 
mm
Note that the threshold drops every 2 metres / 6.5 ft. Refer to the BM100 A 
Service Manual for the threshold value for each 2 metre / 6.5 ft step. 
