KROHNE BM 100A Handbook User Manual
Page 38

38
Handbook BM 100 A
1.6.0 SERIAL I/O
(Serial Input/Output)
For integrating into a digital
network
Select 1200, 2400, 4800,
9600 or 19600 baud
1.6.1 BAUDRATE*
1200 bd
Defines transmission rate of
information between BM100 and
remote terminal. HART/SMART
outputs are fixed at 1200 baud
Enter a number from 0 to
255**
1.6.2 ADDRESS
000
Address of a device in a digital
network. If HART protocol is used,
entering >0 will configure the gauge
for multi-drop networks & output
current 1 to 4 mA.
*RS485 interface only
**0 to 15 if HART.
1.7.0 STRAP. TAB
(Strapping Table)
For calibrating the gauge for
volume measurement.
Select m³, litre, US gal.,
GB, gal., Ft.³, bbl, kg,
metric ton, ton GB or ton
US.
1.7.1 VOL.
UNIT
(Volume / Mass Unit)
m³
Selects the volume or mass unit for a
volume strap table.
Select point 01 to 50,
enter level and then
volume values
respectively.
1.7.2 INPUT.TAB.
(Strapping Table Input)
00
Table or calibrating the gauge point
by point in terms of product level and
volume values.
Select SURE YES or
SURE NO.
1.7.3 DELETE
TAB.
(Strap Table Delete)
SURE NO
Clears the data from the existing
strapping table.