KROHNE BM 100A Handbook User Manual
Page 38

 
 
38
Handbook BM 100 A
 
1.6.0 SERIAL I/O 
(Serial Input/Output)
For integrating into a digital 
network 
Select 1200, 2400, 4800, 
9600 or 19600 baud 
1.6.1 BAUDRATE*
1200 bd
Defines transmission rate of 
information between BM100 and 
remote terminal. HART/SMART 
outputs are fixed at 1200 baud 
Enter a number from 0 to 
255** 
1.6.2 ADDRESS
000
Address of a device in a digital 
network. If HART protocol is used, 
entering >0 will configure the gauge 
for multi-drop networks & output 
current 1 to 4 mA. 
*RS485 interface only 
**0 to 15 if HART. 
 
1.7.0 STRAP. TAB 
(Strapping Table)
For calibrating the gauge for 
volume measurement. 
Select m³, litre, US gal., 
GB, gal., Ft.³, bbl, kg, 
metric ton, ton GB or ton 
US. 
1.7.1 VOL.
UNIT
(Volume / Mass Unit)
m³
Selects the volume or mass unit for a 
volume strap table. 
Select point 01 to 50, 
enter level and then 
volume values 
respectively. 
1.7.2 INPUT.TAB.
(Strapping Table Input)
00
Table or calibrating the gauge point 
by point in terms of product level and 
volume values. 
Select SURE YES or 
SURE NO. 
1.7.3 DELETE
TAB.
(Strap Table Delete)
SURE NO
Clears the data from the existing 
strapping table. 
 
 
