beautypg.com

Technical data – KROHNE BW 25 EN User Manual

Page 46

background image

8

TECHNICAL DATA

46

BW25

www.krohne.com

09/2010 - 4000432302 MA BW25 R05 en

Technical data, indicator M10

M9 ESK2A HART

ESK2A HART® configuration

Manufacturer name (code)

KROHNE Messtechnik (69 = 45h)

Model name

ESK2A (226 = E2h)

HART TM protocol revision

5.9

Device revision

1

Physical layer

FSK

Device category

Transmitter without galvanic isolation

M9 ESK2A process variable

ESK2A process variable, level

Values [%]

Signal output [mA]

Over range

+102.5 (± 1%)

20.24...20.56

Device error identification

> 106.25

> 21.00

Maximum

131.25

25

Multi-drop operation

-

4.5

Min. U

ext.

12VDC

Cable gland

Standard: without

Terminal connection

2,5mm

2

Power supply

24 VDC +/- 30%

Min. power supply for HART TM

18 VDC

Measuring signal

4.00...20.00 mA = 0...100% level value in 2-wire technology

Power supply influence

< 0.1 %

Dependence on external resistance

< 0.1 %

Temperature influence

< 5 µA/K

Max. external resistance / load

≤ 630 Ohm

Min. load for HART

≥ 250 Ohm

Software firmware version

02.14

Ident No:

3209470500

M10 HART

Manufacturer name (code)

KROHNE Messtechnik (69 = 45h)

Model name

M10 (234 = EA)

HART TM protocol revision

5.9

Device revision

1

Physical layer

FSK

Device category

Transmitter

MA_BW25_R05_en_432302_PRT.book Page 46 Tuesday, September 28, 2010 8:17 AM