OPERATION

38

BW25

www.krohne.com

09/2010 - 4000432302 MA BW25 R05 en

3.10

4mA CALIBR.

This menu item allows precise calibration of the current

output. The device generates a fixed analogue output of

4.00 mA. If the measured value differs from the one

displayed, the measured value must be entered. When the

menu is exited, the corrected value is saved.

3.11

20mA CALIBR.

This menu item allows precise calibration of the current

output. The device generates a fixed analog output of 20.00

mA. If the measured value deviates from the displayed one,

then the measured value must be input. When the menu is

exited, the corrected value is saved.

3.12

ALARM

CURRENT

OFF

Measured values > 100% are indicated as a current signal

up to a maximum of 22 mA.

ON

In the event of an error the current output is set to the fixed

value of 22m A.

3.13

END & UNIT

The level meter and end value can be changed.

3.13.1

LEVEL

For a units list, see Section 7.4 of the manual

3.13.2

Inactive

3.14

Inactive

3.15

INP. CODE

YES

The input code is used to prevent authorised adjustment of

the measurement parameters. The input code is not active

by default. If YES is selected, the last code entered must be

typed in. The default code is: → → → ^ ^ ^

↑ ↑ ↑ If, after confirmation with YES, the key → is pressed,

then a new individual, nine-digit code can be typed in. The

display shows the required key combination.

NO

3.16

BASIC

SETTING

YES

This menu item can be used to select the calibrated basic

setting. This can be helpful if operating data have been

changed a number of times. This menu item cannot be

used to reset the calibration.

NO