OPERATION

36

BW25

www.krohne.com

09/2010 - 4000432302 MA BW25 R05 en

6.5.3 Menu explanations

Level Designation

Selection/input
options

Explanation

1.1.1

OUTPUT B1

INACTIVE

LEV.VALUE B1

Level value switching point. A numeric value between 0.0 ...

100% of the level value can be set. If the current level value

exceeds this set switching point, then output B1 is

activated.

Note

Note

Note

Note

The function NC or NO can be selected using menu 3.3.

1.1.2

OUTPUT B1

HYST.B1

Hysteresis setting for the level value switching point. Value

range 0…switching point. Example: If, under 1.1.1, a

switching point of 200 is set, a hysteresis value of 0...200

can be set. If a value of 0 is entered, this output has no

hysteresis. If a value of 20 is entered, the output works as

follows: If the current level value exceeds the value of 200,

the output switches. 3 If the current level value is under

the hysteresis value of 180, the switching output returns to

its normal state. 4

Note

Note

Note

Note

If this function is inverted, the output under menu 3.3 must

be set from NO 1 to NC 2 or vice versa. This function is

not activated with the counter switching point.

1.2.1

OUTPUT B2

INACTIVE

LEV.VALUE B2

See LEV.VALUE B1

1.2.2

OUTPUT B2

HYST.B2

See HYST. B1

1.3

DISPLAY

LEVEL

%

LEVEL & %

1.4

TIME CONST.

Setting : 1 ... 20 seconds

Note

Note

Note

Note

The settable time constant affects the current output and

the displayed current level. It thus allows attenuated

depiction when the display is not calm. If the current level

is polled via HART communication, then the transferred

measured value is dependent on the time constant.