Service, 3 spare parts availability, 4 availability of services – KROHNE OPTISENS PAS 2000 EN User Manual
Page 30
6
SERVICE
30
OPTISENS PAS 2000
www.krohne.com
11/2008 • MA PAS 2000 R04 en
Cleaning the sensor as a slave (along with another sensor)
•
Select the sensor in the main menu by using ↑ or ↓.
•
Press ^ for approximately five seconds to enter the sensor menu.
•
Use ↑ or ↓ to select Cleaning
Cleaning
Cleaning
Cleaning and press ^.
•
The parameters Cleaner
Cleaner
Cleaner
Cleaner, Interval min
Interval min
Interval min
Interval min and Lenght sec
Lenght sec
Lenght sec
Lenght sec in the Cleaning
Cleaning
Cleaning
Cleaning submenu are set for
the sensor being the master.
•
Set Relay
Relay
Relay
Relay to Along #1
Along #1
Along #1
Along #1 or Along #2
Along #2
Along #2
Along #2 depending on what relay the master sensor uses.
•
If needed, specify the extra freeze time in seconds (Freeze sec
Freeze sec
Freeze sec
Freeze sec).
6.3 Spare parts availability
The manufacturer adheres to the basic principle that operational spare parts for each device or
each important accessory part will be kept available for a period of 10 (ten) years after delivery of
the last production run for that device.
Operational spare parts are defined as parts that are subject to faults in normal operation.
6.4 Availability of services
The manufacturer offers a range of services to support the customer after expiration of the
warranty. These include repair, technical support and training.
NOTE!
For more precise information, please contact your local representative.
.book Page 30 Friday, November 28, 2008 2:32 PM