Menu "System"

5.2.2 Menu for PAS 2000 ORP

Menu "Settings"

Menu "Calibrate"

Submenu

Description

Type

Type

Type

Type

Type of sensor, read only.

Serial

Serial

Serial

Serial

Serial number of the sensor, read only.

SoftW

SoftW

SoftW

SoftW

Software version of the sensor, read only.

Press ^ to go to "info" read only menu. This menu is for KROHNE internal use.

Info

Info

Info

Info

MS 1 mV

Millivolt value for the first calibration point.

MS 2 mV

Millivolt value for the second calibration point.

pH 1

pH for the first calibration point.

pH 2

pH for the second calibration point.

RV 0 mV

Raw value for 0 mV.

RV 200 mV

Raw value for 200 mV.

Ch1

Raw value.

MS mV

Electrode output in millivolt.

pH

Current measure value.

Samp/s

Number of samples per second.

Service

Service

Service

Service

Not accessible for users.

Submenu

Description

Tag

Tag

Tag

Tag

Name of the sensor (10 characters) shown in the main display.

I-Time

I-Time

I-Time

I-Time

Integration time or dampening - can be set up to 999 seconds. Normal value is 5-10
seconds.

Analog

Analog

Analog

Analog

"None", "Out1", "Out2", "Out3", "Out4", "Out1+2" or "Out3+4". Pick which analog
output(s) to be used with sensor.

Submenu

Description

Adjust

Adjust

Adjust

Adjust

"No", "Reset" or "Set". Automatic offset adjustment.

Offset

Offset

Offset

Offset

Offset for the electrode.

@Buffer

@Buffer

@Buffer

@Buffer

Buffer used for adjustment, saved during the automatic offset adjustment.

ORP mV

ORP mV

ORP mV

ORP mV

Current reading.

New electr.

New electr.

New electr.

New electr.

Date when electrode was last changed. Push ^ to edit the date, then ^ again to store it.