Operation – KROHNE OPTISENS PAS 2000 EN User Manual
Page 23
OPERATION
5
23
OPTISENS PAS 2000
www.krohne.com
11/2008 • MA PAS 2000 R04 en
Menu "System"
5.2.2 Menu for PAS 2000 ORP
Menu "Settings"
Menu "Calibrate"
Submenu
Description
Type
Type
Type
Type
Type of sensor, read only.
Serial
Serial
Serial
Serial
Serial number of the sensor, read only.
SoftW
SoftW
SoftW
SoftW
Software version of the sensor, read only.
Press ^ to go to "info" read only menu. This menu is for KROHNE internal use.
Info
Info
Info
Info
MS 1 mV
Millivolt value for the first calibration point.
MS 2 mV
Millivolt value for the second calibration point.
pH 1
pH for the first calibration point.
pH 2
pH for the second calibration point.
RV 0 mV
Raw value for 0 mV.
RV 200 mV
Raw value for 200 mV.
Ch1
Raw value.
MS mV
Electrode output in millivolt.
pH
Current measure value.
Samp/s
Number of samples per second.
Service
Service
Service
Service
Not accessible for users.
Submenu
Description
Tag
Tag
Tag
Tag
Name of the sensor (10 characters) shown in the main display.
I-Time
I-Time
I-Time
I-Time
Integration time or dampening - can be set up to 999 seconds. Normal value is 5-10
seconds.
Analog
Analog
Analog
Analog
"None", "Out1", "Out2", "Out3", "Out4", "Out1+2" or "Out3+4". Pick which analog
output(s) to be used with sensor.
Submenu
Description
Adjust
Adjust
Adjust
Adjust
"No", "Reset" or "Set". Automatic offset adjustment.
Offset
Offset
Offset
Offset
Offset for the electrode.
@Buffer
@Buffer
@Buffer
@Buffer
Buffer used for adjustment, saved during the automatic offset adjustment.
ORP mV
ORP mV
ORP mV
ORP mV
Current reading.
New electr.
New electr.
New electr.
New electr.
Date when electrode was last changed. Push ^ to edit the date, then ^ again to store it.
.book Page 23 Friday, November 28, 2008 2:32 PM