3.3.4 Assembly of PAS 2000 to MAA 2000 slide rail immersion holder
As an alternative way of mounting the sensor, an adjustable slide rail holder is available.
Mounting instructions:
•
Mount the slide rail assembly to the side wall of the basin or open channel using the two
predrilled holes. The adjustable limit stop should be on the bottom and the two sliding clamps
above.
•
Take the two clamps off from the slide rail and mount them around the sensor housing. Please
make sure the two guide tracks line-up in one straight line to each other. See figure 3-4!
•
Slide the sensor with the two clamps into the slide rail and make sure that the guide tracks of
the two clamps are properly seated.
•
Adjust the sensor position as necessary (see chapter 3.3.2.) and fasten the limit stop.
1
Slide rail
2
Sensor
3
28 mm / 1.10" clamp
NOTE!
Please make sure that the assembly is mounted in a vertical position. In some applications it
may be desirable to mount at a certain angle. In this case the angle from the horizontal plane
should be greater than or equal to 45°.