5.1 Sensor display
By simultaneously pressing ↓ and ^ you alter between MAC 080 main menu and the sensor
information display for the selected sensor. PAS 2000 has two information pages. The first one
shows in addition to the measured value maximum, minimum and average values for the last
24 hours. The values are based on hour values which means they may jump when a new hour
starts and the oldest is skipped.
The second information display shows the date of the last calibration (PAS 2000 pH only) and the
date the electrode was changed.
5.2 Menu for sensors
Use ↑ or ↓ to select the sensor in the main display. The menu for the selected sensor is accessed
by pressing ^ for five seconds. If the selected sensor is not active (the text No transmitter
No transmitter
No transmitter
No transmitter is
shown) a warning is displayed that asks you to make another choice in order to show the sensor
menu.
5.2.1 Menu for PAS 2000 pH
Menu "Settings"
Menu "Calibrate"
Submenu
Description
Tag
Tag
Tag
Tag
Name of the sensor (10 characters) shown in the main display.
I-Time
I-Time
I-Time
I-Time
Integration time or dampening - can be set up to 999 seconds. Normal value is 5-10
seconds.
Decimals
Decimals
Decimals
Decimals
"1 or "2", number of decimals to show the measured value (only pH).
Reset day
Reset day
Reset day
Reset day
"Yes" or "No", Yes resets the day calculation of min- max- and average.
Analog
Analog
Analog
Analog
"None" , "Out1", "Out2", "Out3", "Out4", "Out1+2", or "Out3+4". Pick which analog
output(s) to be used with sensor.
Submenu
Description
Take sample
Take sample
Take sample
Take sample
"No", "Smart", or "manual". Refer to chapter 10 for details.
Cal Temp
Cal Temp
Cal Temp
Cal Temp
Temperature at which calibration was done.
Measure Temp
Measure Temp
Measure Temp
Measure Temp
Temperature of measured media.
Slope %
Slope %
Slope %
Slope %
Shows the current condition of the electrode. A value between 93 % and 102 % is
acceptable.
Sample 1 pH
Sample 1 pH
Sample 1 pH
Sample 1 pH
Actual pH for the first buffer.
Sample 2 pH
Sample 2 pH
Sample 2 pH
Sample 2 pH
Actual pH for the second buffer.
Calibrated
Calibrated
Calibrated
Calibrated
Date of last air calibration.
New electr.
New electr.
New electr.
New electr.
Date when electrode was last changed. Push ^ to edit the date, then ^ again to store
it.