INSTALLATION
3
13
OPTISENS PAS 2000
www.krohne.com
11/2008 • MA PAS 2000 R04 en
3.3 Installing the sensor
3.3.1 Basic installation instructions
The electrode should be placed deep enough so that it always stays in the measuring media,
even when the level varies (see figure 3-1). The measuring ability of the electrode is influenced
negatively if the column of liquid to be measured is too large. For optimal results the column
should not exceed 300 mm / 12".
The sensor should be mounted in a vertical position. In some applications it may be desirable to
mount at a certain angle. In this case the angle from the horizontal plane should be greater than
or equal to 45°.
3.3.2 Mounting to MAA 2000 telescopic rod immersion holder
The mounting bracket of the telescopic rod is mounted to a handrail or a separate holder.
In case a handrail is not available, a mounting post with a vertical bar for sensor mounting can be
purchased from the manufacturer.
NOTE!
To protect the sensor from damage, the sensor should never be fully submerged into water.
Therefore a maximal immersion depth of 800 mm / 31" has to be maintained.
Figure 3-1:
Installing the PAS 2000 in a flume