ELECTRICAL CONNECTIONS

4

23

TIDALFLUX 4300 F

www.krohne.com

06/2010 - 4000970001 - TIDALFLUX 4300 F R01 en

4.9 Preparing field current cable C, connection to measuring sensor

• The field current cable is not part of the scope of supply.
• The shield is connected in the terminal compartment of the converter directly via the shield

and a clip.

• The shield is connected in the sensor via the special cable gland.
• Bending radius: ≥ 50 mm / 2"

Required materials

• Shielded 2-wire insulated copper cable
• Insulating tubing, size according to the cable being used
• Heat-shrinkable tubing
• DIN 46 228 wire end ferrules: size according to the cable being used

1 Strip the conductor to dimension a.
2 Trim the outer shield to dimension b and pull it over the outer sheath.
3 Crimp wire end ferrules onto both conductors.
4 Pull a shrinkable tube over the prepared cable.

INFORMATION!
Assembly materials and tools are not part of the delivery. Use the assembly materials and tools
in compliance with the applicable occupational health and safety directives.

Figure 4-8: Preparation of field current cable C

a = 50 mm / 2"
b = 10 mm / 0.4"