3.4.6 Control valve

3.4.7 Slope

3.4.8 Mounting advice for difficult situations

If you can not meet the installation conditions install the flowmeter between two containers. The
inlet to the flowmeter must be higher than the outlet of the fluid. In this way you will have a calm
flow into the flowmeter, resulting in a highly accurate measurement. The sizes of the containers
must be proportional to the size of the flowmeter.

Figure 3-7: Installation before control valve

CAUTION!
The accuracy is influenced by the slope. Stay within

±

1% to get the most accurate

measurements!

Figure 3-8: Recommended slope

Figure 3-9: Installing in difficult situations

1 Use a container 2 if the Inlet pipe has a slope > 1%. Make sure that the outlet level of this pipe is below the inlet to the

flowmeter.

2 Inlet container
3 Inlet section of 10 DN
4 Outlet section of 5 DN
5 Outlet container advisable if outlet pipe has a slope > 1%.