beautypg.com

Electrical connections – KROHNE TIDALFLUX 4300 F EN User Manual

Page 20

background image

4

ELECTRICAL CONNECTIONS

20

TIDALFLUX 4300 F

www.krohne.com

06/2010 - 4000970001 - TIDALFLUX 4300 F R01 en

4.6 Prepare signal cable A, connect to measuring sensor

Required materials

• PVC insulation tubing, Ø2.0...2.5 mm / 0.08...0.1"
• Heat-shrinkable tubing
• Wire end ferrule to DIN 46 228: E 1.5-8 for the twisted stranded drain wires (1) and (6)
• 2x wire end ferrules to DIN 46 228: E 0.5-8 for the insulated conductors (2, 3)

1 Strip the conductor to dimension a.
2 Cut the outer shields (60) and (10). Make sure not to damage the stranded drain wires (1) and

(6).

3 Twist the stranded drain wires (6) of the outer shield and the drain wire (1) of the inner shield

(10).

4 Slide an insulating tube over the stranded drain wires (1) and (6).
5 Crimp the wire end ferrules onto conductors 2 and 3 and the stranded drain wires (1) and (6).
6 Pull the heat-shrinkable tubing over the prepared signal cable.

INFORMATION!
Assembly materials and tools are not part of the delivery. Use the assembly materials and tools
in compliance with the applicable occupational health and safety directives.

Figure 4-5: Prepare signal cable A, connect to measuring sensor

a = 50 mm / 2"