7
TECHNICAL DATA
50
TT 51 SERIES
www.krohne.com
09/2010 - 4000754201 - MA TT 51 R01 en
Accuracies in °F
Input type
Temp. range
Min. span
Accuracy
Temp. influence
(Dev. from ref. temp. 68°F)
[°F]
[°F]
[°F]
RTD Pt100
-328...+1562
50
±
0.18°F or ±0.05% of span
±
0.005% of span per °F
RTD PtX
1
Corresp. to max.
4 kΩ
50
±
0.18°F or ±0.05% of span
±
0.005% of span per °F
RTD Ni100
-76...+482
50
±
0.18°F or ±0.05% of span
±
0.005% of span per °F
RTD Ni120
-76...+482
50
±
0.18°F or ±0.05% of span
±
0.005% of span per °F
RTD Ni1000
-58...+356
50
±
0.18°F or ±0.05% of span
±
0.005% of span per °F
RTD Cu10
-58...+392
181
±
2.7°F or ±0.1% of span
±
0.01% of span per °F
T/C type B
+752...+3272
1292
±
1.8°F or ±0.1% of span
2
±
0.005% of span per °F
T/C type C
+32...+4199
392
±
1.8°F or ±0.1% of span
2
±
0.005% of span per °F
T/C type D
+32...+4199
392
±
1.8°F or ±0.1% of span
2
±
0.005% of span per °F
T/C type E
-328...+1832
122
±
0.45°F or ±0.1% of span
2
±
0.005% of span per °F
T/C type J
-328...+1832
122
±
0.45°F or ±0.1% of span
2
±
0.005% of span per °F
T/C type K
-328...+2462
122
±
0.45°F or ±0.1% of span
2
±
0.005% of span per °F
T/C type N
-148...+2372
212
±
0.45°F or ±0.1% of span
2
±
0.005% of span per °F
T/C type N
-418...+212
212
±
1.8°F
2
±
0.05% of span per °F
T/C type R
-58...+3182
572
±
1.8°F or ±0.1% of span
2
±
0.005% of span per °F
T/C type S
-58...+3182
572
±
1.8°F or ±0.1% of span
2
±
0.005% of span per °F
T/C type T
-328...+752
122
±
0.45°F or ±0.1% of span
2
±
0.005% of span per °F
1 (10 ≤ X ≤ 1000)
2 CJC error is not included