beautypg.com

Technical data – KROHNE OPTITEMP TT 51 EN User Manual

Page 44

background image

7

TECHNICAL DATA

44

TT 51 SERIES

www.krohne.com

09/2010 - 4000754201 - MA TT 51 R01 en

Output

Output

Output

Output

Output signal

4...20 mA, 20...4 mA or customized

Temperature linear for RTD & T/C

HART

®

protocol

HART

®

6

HART

®

physical layer

FSK 1200

Representation

T1 or T2 or difference, average, min. or max. of T1 and T2

Update time

Single input: ≈300 ms; Double input: ≈600 ms

Adjustable output filtering

0...60 s (time constant)

Permissible load

635 Ω at 24 VDC incl. 250 Ω loop resistance

NAMUR compliance

Current limitations and failure currents acc. to NAMUR NE 43

Configuration

Configuration

Configuration

Configuration

ConSoft

The PC configuration software, ConSoft, is a versatile and user-friendly tool

for transmitter configuration, loop check-up and sensor diagnostics.

It runs on Windows 2000, XP and Vista.

ConSoft is part of the complete configuration kit ICON, which also contains a

USB Interface and necessary cables.

Alternatives

Hand held communicator, e.g. FC375/FC475 (Emerson)

Management systems, e.g. AMS (Emerson) and PDM (Siemens)

EDD enabled systems

DTM/FDT enabled systems

Approvals and certifications

CE

The device fulfils the statutory requirements of the EC directives.

The manufacturer certifies that these requirements have been met by

applying the CE marking.

Ex approvals

Ex approvals

Ex approvals

Ex approvals

Standard version

Without

Non-incendive version

ATEX: II 3 G Ex nL IIC T4...T6

Intrinsically safe (IS) version

ATEX: II 1 G Ex ia IIC T4...T6 (only in-head transmitter)

ATEX: II 2(1) G Ex ia IIC T4...T6 (only rail-mount transmitter)

Others

In preparation: FM, CSA, IECEx and GOST

Other standards and approvals

Other standards and approvals

Other standards and approvals

Other standards and approvals

Electromagnetic compatibility

Directive: 2004/108/EC

Harmonized standards: EN 61326-1:2006; EN 61326-3-1:2009;

NAMUR NE 21

Immunity performance: criteria A; surge test influence max. ±0.5% of span

Vibration resistance

Acc. to IEC 60068-2-6, test Fc, 10...2000 Hz, 10 g

Shock resistance

Acc. to IEC 60068-2-31, test Ec